NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded sensors, ... — FDA record Z-2114-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 19, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded sensors, a paired Android tablet, and proprietary software, packaged in a portable field case. Eye-tracking headset system is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects. The EYE-SYNC is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI).
Reason
Eye syncing issues causing the device to be inoperable. Unable to be repaired due to discontinuation of support for the Pico 2 platform, which was no longer in production or serviceable as of August 31, 2023.
Classification
Class II
Recalling firm
NeuroSync, Inc.
Quantity
27 units
Agency-reported status
Ongoing
Source record identifier
Z-2114-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel No X02. UDI-DI: 00868096000300. Serial Numbers: PA7D50NGF4020311G (Lot P-10114-13), PA7D50NGF4020191G (Lot P-10114-13), PA7D40NGEA080237G (Lot P-10114-12) Software Revisions: Software version uniform across all affected devices

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

United States Nationwide distribution in the states of New York (NY), Texas (TX), and Georgia (GA).

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 23, 2025
Recall initiation date
Sep 19, 2023
Center classification date
Jul 11, 2025