Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac). — FDA record Z-2110-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac).
Reason
There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used.
Classification
Class I
Recalling firm
Abbott
Quantity
675 units
Agency-reported status
Ongoing
Source record identifier
Z-2110-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 05415067036025, Batch Numbers: 8597228, 8598983, 8601676, 8611443, 8619197, 8627877, 8650955, 8663679, 8795668, 8815423, 8824607, 8830530, 8846835

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Slovakia, Czech Republic.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 19, 2023
Recall initiation date
Jun 12, 2023
Center classification date
Jul 11, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records