The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009 — FDA record Z-2104-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009
Reason
A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.
Classification
Class I
Recalling firm
Baxter Healthcare Corporation
Quantity
19861 units
Agency-reported status
Ongoing
Source record identifier
Z-2104-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 00085412610900, Software Version v9.02.01, All Serial Numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 19, 2023
Recall initiation date
Jun 15, 2023
Center classification date
Jul 13, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.