Medline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A; 2) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A... — FDA record Z-2102-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 7, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Medline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A; 2) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 3) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 4) TOTE SB NGUYEN CAPSULOTOMY, Model Number: DYKM1756A; 5) TOTE MOMENI SB CAPSULOTOMY REC, Model Number: DYKM1765A; 6) TOTE SAFRAN OSC HIP ARTHRSCPY, Model Number: DYKM1767B; 7) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 8) TOTE CURTIN OSC HAND REPAIR, Model Number: DYKM1781A; 9) TOTE FOX OSC HAND REPAIR, Model Number: DYKM1783A; 10) TOTE YAO OSC HAND REPAIR, Model Number: DYKM1796A; 11) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 12) TOTE YAO OSC WRIST ARTHROSCOPY, Model Number: DYKM1807; 13) TOTE CARROLL BLOCK BLOOD PATCH, Model Number: DYKM1864; 14) TOTE SB STONE JWIRE LUMPECTOMY/SEN NODE, Model Number: DYKM1925; 15) TOTE SB NGUYEN LIPOSUCTION, Model Number: DYKM1927A; 16) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977B; 17) CHEST TUBE SUPPLIES, Model Number: DYKM2131
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class II
Recalling firm
Medline Industries, LP
Quantity
1516 units
Agency-reported status
Ongoing
Source record identifier
Z-2102-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1) CVI4500A, UDI-DI: 10653160327646(each), 653160327649(case), Lot Number: 2020121550; 2) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024030690; 3) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024052290; 4) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024060590; 5) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2025012890; 6) DYKM1751A, UDI-DI: 10193489111002(each), 20193489111009(case), Lot Number: 2023121190; 7) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023060290; 8) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023070590; 9) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023111590; 10) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023120590; 11) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024022190; 12) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024080190; 13) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024101490; 14) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2023050490; 15) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2023071290; 16) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2024012490; 17) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2024071790; 18) DYKM1767B, UDI-DI: 10193489112368(each), 20193489112365(case), Lot Number: 2023102790; 19) DYKM1776, UDI-DI: 10888277733794(each), 20888277733791(case), Lot Number: 2024031590; 20) DYKM1776, UDI-DI: 10888277733794(each), 20888277733791(case), Lot Number: 2025013190; 21) DYKM1781A, UDI-DI: 10653160352730(each), 653160352733(case), Lot Number: 2024092390; 22) DYKM1783A, UDI-DI: 10193489111330(each), 20193489111337(case), Lot Number: 2024053190; 23) DYKM1783A, UDI-DI: 10193489111330(each), 20193489111337(case), Lot Number: 2024062890; 24) DYKM1783A, UDI-DI: 10193489111330(each), 20193489111337(case), Lot Number: 2025011090; 25) DYKM1796A, UDI-DI: 10653160354888(each), 653160354881(case), Lot Number: 2024041790; 26) DYKM1796A, UDI-DI: 10653160354888(each), 653160354881(case), Lot Number: 2024061890; 27) DYKM1796A, UDI-DI: 10653160354888(each), 653160354881(case), Lot Number: 2024072390; 28) DYKM1796A, UDI-DI: 10653160354888(each), 653160354881(case), Lot Number: 2024093090; 29) DYKM1796A, UDI-DI: 10653160354888(each), 653160354881(case), Lot Number: 2025010890; 30) DYKM1796A, UDI-DI: 10653160354888(each), 653160354881(case), Lot Number: 2025020690; 31) DYKM1801B, UDI-DI: 10653160360421(each), 653160360424(case), Lot Number: 2024011790; 32) DYKM1801B, UDI-DI: 10653160360421(each), 653160360424(case), Lot Number: 2024062890; 33) DYKM1801B, UDI-DI: 10653160360421(each), 653160360424(case), Lot Number: 2024100790; 34) DYKM1807, UDI-DI: 10193489191219(each), 20193489191216(case), Lot Number: 2024032990; 35) DYKM1807, UDI-DI: 10193489191219(each), 20193489191216(case), Lot Number: 2024071990; 36) DYKM1864, UDI-DI: 10193489191257(each), 20193489191254(case), Lot Number: 2023101680; 37) DYKM1864, UDI-DI: 10193489191257(each), 20193489191254(case), Lot Number: 2024010880; 38) DYKM1925, UDI-DI: 10193489190311(each), 20193489190318(case), Lot Number: 2023112780; 39) DYKM1925, UDI-DI: 10193489190311(each), 20193489190318(case), Lot Number: 2024040180; 40) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023072590; 41) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023080190; 42) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023081490; 43) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023121490; 44) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2024012590; 45) DYKM1977B, UDI-DI: 10653160360797(each), 653160360790(case), Lot Number: 2023063090; 46) DYKM1977B, UDI-DI: 10653160360797(each), 653160360790(case), Lot Number: 2023120890; 47) DYKM1977B, UDI-DI: 10653160360797(each), 653160360790(case), Lot Number: 2024091290; 48) DYKM2131, UDI-DI: 10195327058715(each), 40195327058716(case), Lot Number: 23ALA008

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 13, 2026
Recall initiation date
Jan 7, 2026
Center classification date
May 7, 2026