NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and... — FDA record Z-2093-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs in transport medium and bronchoalveolar lavage (BAL) specimens from individuals suspected of COVID-19 by their healthcare provider (HCP). REF Number: 300800
Reason
Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.
Classification
Class III
Recalling firm
NeuMoDx Molecular Inc
Quantity
1,086 test strips
Agency-reported status
Ongoing
Source record identifier
Z-2093-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 10814278021264 Lot Number: 123909 Exp. Date: 08-Dec-2023

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of FL, MI, NM, NY, TN, TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 12, 2023
Recall initiation date
May 15, 2023
Center classification date
Jul 5, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.