Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software V... — FDA record Z-2081-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 14, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No
Reason
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
Classification
Class I
Recalling firm
Edwards Lifesciences, LLC
Quantity
13240
Agency-reported status
Ongoing
Source record identifier
Z-2081-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Model No: OPTI18; UDI: 00690103180565; All lots up to: BSLC9155 Serial Numbers: 130451 130466 130472 130593 130709 130764 130818 130821 130847 130948 130952 131038 131043 131047 131182 131193 131231 131236 131289 131311 131317 131433 131435 131513 131516 131553 131554 131606 131623 131625 131628 131682 131737 131909 131929 131966 132175 132201 132366 132395 132416 132430 132435 132505 132832 132958 132959 BSLC0110 BSLC0166 BSLC0661 BSLC0661 BSLC0826 BSLC0826 BSLC0826 BSLC0826 BSLC0855 BSLC0855 BSLC0887 BSLC0887 BSLC0979 BSLC1114 BSLC1114 BSLC1211 BSLC1308 BSLC1469 BSLC1656 BSLC1660 BSLC1815 BSLC3221 BSLC4153 BSLC4400 BSLC4461 BSLC4513 BSLC5027 BSLC5027 BSLC5320 BSLC5367 BSLC5765 BSLC6421 BSLC6467 BSLC7335 BSLC7514 BSLC7515 BSLC8175 BSLC8385 BSLC8535 BSLC8657 BSLC9100 BSLC9104 BSLC9105 BSLC9152 BSLC9155

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 16, 2025
Recall initiation date
May 14, 2025
Center classification date
Jul 10, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.