What the record says
- Product
- BETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number: 110045160
- Reason
- Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.
- Classification
- Class II
- Recalling firm
- T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
- Quantity
- N/A
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2061-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 10818674025772 Lot Numbers: 21E28, 21R15, 21R16, 21R32, 21R35, 22C19, 22E29, 22F10, 22F35, 22J07, 22J08, 22J09, 22J18, 22J19, 22J27, 22J28, 22J33, 22K23, 22K24, 22K32, 22K33, 22K38, 22M03, 22M04, 22M24, 22P03, 22P09, 22P10, 22P11, 22P13, 22P14, 22P16 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Japan, Netherlands.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jul 12, 2023
- Recall initiation date
- May 4, 2023
- Center classification date
- Jul 3, 2023