TruSignal Integrated Ear Sensor with Datex Connector, REF TS-E4-GE; Monitor, physiological, patient(with arrhythmia d... — FDA record Z-2032-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date May 19, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- TruSignal Integrated Ear Sensor with Datex Connector, REF TS-E4-GE; Monitor, physiological, patient(with arrhythmia detection or alarms)
- Reason
- There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.
- Classification
- Class I
- Recalling firm
- GE Healthcare Finland Oy
- Quantity
- 2707 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2032-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | GTIN 00840682103428 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Jul 12, 2023
- Recall initiation date
- May 19, 2023
- Center classification date
- Jul 6, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.