Recall archive / FDA device enforcement
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MICRODISECTOMY PACK, Medline... — FDA record Z-2030-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 19, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585B; 2. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585C; 3, MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585D.
- Reason
- During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
- Classification
- Class II
- Recalling firm
- Medline Industries, LP
- Quantity
- 924 kits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2030-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Medline Kit Number/SKU DYNJ80585B: UDI/DI each 10195327211158, UDI/DI case 40195327211159, Lot Number: 22GBV101; Medline Kit Number/SKU DYNJ80585B: UDI/DI each 10195327211158, UDI/DI case 40195327211159, Lot Number: 24IBG473; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25EBT656; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25DBN421; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25CBN640; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25BBL172; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25BBD651; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25ABA455; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 24KBO858; Medline Kit Number/SKU DYNJ80585C: UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 24JBH693; Medline Kit Number/SKU DYNJ80585D: UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 26ABH754; Medline Kit Number/SKU DYNJ80585D: UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25LBA242; Medline Kit Number/SKU DYNJ80585D: UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25HBT362; Medline Kit Number/SKU DYNJ80585D: UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25GBL430; Medline Kit Number/SKU DYNJ80585D: UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25FBT402. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- May 6, 2026
- Recall initiation date
- Mar 19, 2026
- Center classification date
- Apr 30, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.