OES 4000 Hysteroscope, Model/Catalog Number: A4674A — FDA record Z-2029-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 18, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
OES 4000 Hysteroscope, Model/Catalog Number: A4674A
Reason
The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
Classification
Class II
Recalling firm
Olympus Corporation of the Americas
Quantity
1 unit
Agency-reported status
Ongoing
Source record identifier
Z-2029-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of NY.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 12, 2024
Recall initiation date
Apr 18, 2024
Center classification date
Jun 6, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records