Oxylog 3000 Plus emergency and transport ventilator — FDA record Z-2018-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Oxylog 3000 Plus emergency and transport ventilator
Reason
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Classification
Class I
Recalling firm
Draeger Medical, Inc.
Quantity
300 units
Agency-reported status
Ongoing
Source record identifier
Z-2018-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog No. 5704811 & 5704813; UDI-DI: 04048675398295; All Serial No.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Faeroe, Finland, France, French Polynesia, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Macau, Malawi, Malaysia, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, North Korea, Norway, Oman, Pakistan, Panama, Pap. New Guinea, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Rwanda, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Sint Maarten, Slovenia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, Utd. Arab. Emirates, Uzbekistan, Vietnam, Yemen, Zambia, and Zimbabwe.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 19, 2023
Recall initiation date
Jun 12, 2023
Center classification date
Jul 7, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records