Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStatio... — FDA record Z-2017-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 8, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main 10885403512629 [303], BD Pyxis MedStation ES Tower 10885403512674 [352], BD Pyxis Anesthesia Station ES 10885403477836 [327], BD Pyxis MedStation ES (Main) 10885403512667 [323], MD BD Pyxis SupplyStation V9 [317], Security Module [139698-01], BD Knowledge Portal for Medication Technologies [136607-01], STOCKSTN V10.X WRD/WL SCANNERS [132-53-02], PRS STOCKSTN V10.X WRD/WL SCANNERS [132-53-01], STOCKSTN V10.X WIRED SCANNER [132-52-02], PRS STOCKSTN V10.X WIRED SCANNER [132-52-01], BD Pyxis CII Safe ES Desktop PC [1156-00], BD Pyxis SupplyStation Panel PC V11 [1132-00], BD Pyxis SupplyRoller V11 [1127-00], Dell 640 Server [1128-00], BD Pyxis SupplyStation System V11 [1123-00], BD Pyxis CII Safe ES 10885403512605 [1116-00], BD Pyxis Enterprise Server 10885403518348, 10885403520341 [1115-00], BD Pyxis SupplyRoller V10 [1101-00], BD Pyxis CIISafe V9.X SGL IM BIO 10885403512520 [111-221], BD Pyxis CIISafe V7.X DBL INTG MAIN BIO, SOLID DO [111-194], BD Pyxis CIISafe V8.X Desktop PC, Bio [107-245-01], BD Pyxis SupplyStation V10 [347], BD Pyxis CII Safe V9.X DESKTOP PC [107-252-01], BD Pyxis CII Safe V7.X DBL INTG MAIN BIO CLEAR DO [111-191], BD Pyxis CII Safe V7.XSGL INTG MAIN BIO SOLID DO [111-198], BD Pyxis CII SafeV8.XDBL INTG MAIN BIO SOLID DO [111-202-01], BD Pyxis CII SafeV8.XSGL INTG MAIN BIO SOLID DO [111-206-01], BD Pyxis CII Safe V9.X DBL IM BIO SLD DOORS [111-215], BD Pyxis CII Safe V9.X SGL IM BIO CLR DOORS [111-216], BD Pyxis CII Safe V9.X DBL IM BIO CLR DOORS [111-219], Supply itParActive RF V11 (Supply Kanban) [138566-01]
Reason
During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.
Classification
Class II
Recalling firm
CareFusion 303, Inc.
Quantity
6,332
Agency-reported status
Ongoing
Source record identifier
Z-2017-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityEach customer was provided a unique list of their affected devices that included computer names and atlas keys. Distributed: April 14, 2014 - March 21, 2025 REF or Name [UDI-DI]: Supply Server [N/A], CCE Enterprise SW Site License (2-5) [N/A], 806 [N/A], 303 [10885403512629], 352 [10885403512674], 327 [10885403477836], 323 [10885403512667], 317 [N/A], 139698-01 [N/A], 136607-01 [N/A], 132-53-02 [N/A], 132-53-01 [N/A], 132-52-02 [N/A], 132-52-01[N/A] , 1156-00 [N/A], 1132-00 [N/A], 1127-0 [N/A], 1128-00 [N/A], 1123-00 [N/A], 1116-00 [N/A], 1115-00 [10885403518348, 10885403520341], 1101-00 [N/A], 111-221 [10885403512520], 111-194 [N/A], 107-245-01 [N/A], 347[N/A], 107-252-01 [N/A], 111-191 [N/A], 111-198 [N/A], 111-202-01 [N/A], 111-206-01 [N/A], 111-215 [10885403512520], 111-216 [10885403512520], 111-219 [10885403512520], 138566-01 [N/A]

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of PR, DE, GU, MO, IA, IL, CA, MD, MA, VA, PA, NY, WA, NJ, AZ, CO, OH, FL, WI, TX, CT, LA, MI, WY, MN, AR, KY, OR, NM, TN, NC, SC, IN, VT, KS, AL, GA, WV, NV, ID, DC, MS, NE, HI, OK, UT, ND, MT, NH, SD, ME, RI, AK, MP and the countries of SA, CA, NZ , CH, FR, IT, BE, AU, BH, QA, AE, GB, EG, AR, JO, DE, BM, TH, LB, BR, IE, FI, MX, TW, ID, SG, BS, IN, MA, GF, ES, HK.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 2, 2025
Recall initiation date
May 8, 2025
Center classification date
Jun 24, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records