Recall archive / FDA device enforcement
Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Mod... — FDA record Z-1981-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 19, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
- Reason
- Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
- Classification
- Class I
- Recalling firm
- American Contract Systems Inc
- Quantity
- 7,311 kits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1981-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US distribution to MN. No OUS distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- May 13, 2026
- Recall initiation date
- Mar 19, 2026
- Center classification date
- May 6, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.