GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S — FDA record Z-1980-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 8, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S
Reason
Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.
Classification
Class II
Recalling firm
Ohio Medical Corporation
Quantity
36 units
Agency-reported status
Ongoing
Source record identifier
Z-1980-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL317087, ZL317088, ZL317090, ZL322386, ZL322387, ZL322390, ZL322392, ZL322409, ZL322412, ZL322439, ZL322441, ZL322442, ZL322443, ZL322445, ZL322447, ZL322451, ZL322455, ZL322456, ZL322462, ZL322463, ZL322466, ZL322467, ZL322470, ZL322472, ZL322473, ZL322837.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 25, 2025
Recall initiation date
May 8, 2025
Center classification date
Jun 13, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.