V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS). — FDA record Z-1980-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 2, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).
Reason
The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.
Classification
Class II
Recalling firm
SenTec AG
Quantity
11 in the U.S.
Agency-reported status
Ongoing
Source record identifier
Z-1980-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 28, 2023
Recall initiation date
May 2, 2023
Center classification date
Jun 20, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.