Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEU... — FDA record Z-1974-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 25, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.
Reason
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
Classification
Class I
Recalling firm
QUIDEL CARDIOVASCULAR INC
Quantity
16,194
Agency-reported status
Ongoing
Source record identifier
Z-1974-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Lots(Expiration): 97000HS/30014613337911/T13666N(2023-06-26), T13667N(2023-07-03), T13669N(2023-07-08), T13706N(2023-07-21), T13826N(2023-09-03), T13944N(2023-10-14), T13948N(2023-10-20), T13949N(2023-10-22), T14019N(2023-11-05), T14020N(2023-11-05), T14023N(2023-11-11); 97021HS/30014613337935/T13665RN(2023-06-25), T13707RN(2023-06-28), T13709RN(2023-08-11), T13827RN(2023-09-04), T13829RN(2023-09-18), T13942RN(2023-10-09), T13950RN(2023-10-23); 97022HS/30014613337942/T13664RN(2023-06-20), T13708RN(2023-08-03), T13943RN(2023-10-10); 97000HSEU/30014613337928/T13668RBN(2023-07-05), T13705RN(2023-07-14), T13765RBN(2023-08-26), T13825RBN(2023-08-27), T13828RBN(2023-09-18), T13946RBN(2023-10-16); 97000HSEUJP/T13668RN(2023-07-05), T13765RN(2023-08-26), T13828RN(2023-09-18), T13946RN(2023-10-16); 97000HZ01/T13763(2023-08-19), T13764(2023-08-21), T13830(2023-09-24), T13831(2023-09-26), T13940(2023-10-01), T13941(2023-10-03), T13947(2023-10-21), T13951(2023-10-27), T14021(2023-11-06), T14022(2023-11-10), T14024(2023-11-15), T14025(2023-11-19), T14041(2023-12-13), T14042(2023-12-18), T14043(2023-12-19), T14044(2023-12-20), T14045(2023-12-24); 97000QIL/T13825RNQ(2023-08-27). 97000HS/97021HS/97022HS:Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24). 97000HS/97000HSEU/97000QIL/97021HS: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN, T13831NQ, T13831RBNQ, T13831RNQ(2023-09-26); T14021N, T14021RBN, T14021RN, T14021NQ, T14021RBNQ, T14021RNQ(2023-11-06); T14022N, T14022RBN, T14022RN, T14022NQ, T14022RBNQ, T14022RNQ(2023-11-10); T14024N, T14024RNB, T14024RN, T14024NQ, T14024RNBQ, T14024RNQ(2023-11-15); T14025N, T14025RBN, T14025RN, T14025NQ, T14025RBNQ, T14025RNQ(2023-11-19); T14041N, T14041RBN, T14041RN, T14041NQ, T14041RBNQ, T14041RNQ(2023-12-13); T14042N, T14042RBN, T14042RN, T14042NQ, T14042RBNQ, T14042RNQ(2023-12-18); T14043N, T14043RBN, T14043RN, T14043NQ, T14043RBNQ, T14043RNQ(2023-12-19); T14044N, T14044RBN, T14044RN, T14044NQ, T14044RBNQ, T14044RNQ(2023-12-20); T14045N, T14045RBN, T14045RN, T14045NQ, T14045RBNQ, T14045RNQ(2023-12-24). 97000HSEU/97000HSEUJP: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24).

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 12, 2023
Recall initiation date
May 25, 2023
Center classification date
Jul 6, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.