Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect... — FDA record Z-1972-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual
Reason
Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.
Classification
Class II
Recalling firm
Boston Scientific Neuromodulation Corporation
Quantity
21,533
Agency-reported status
Ongoing
Source record identifier
Z-1972-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityDriver UDI-DI: 00884662000574, UPN: 102-9800, All Lots. Superion Indirect Decompression System IFU (92479815-02), Superion IDS Kit IFU (92479820-02), Surgical Technique Manual (92479821-02)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 21, 2023
Recall initiation date
May 12, 2023
Center classification date
Jun 15, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records