Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Opti... — FDA record Z-1971-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 16, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
Reason
When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
Classification
Class II
Recalling firm
Ellex Medical Pty Ltd.
Quantity
57 systems in total (6 units shipped to United States)
Agency-reported status
Ongoing
Source record identifier
Z-1971-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Nos., UDI IPY0374, (01) 0 9342395 00012 0 ISRY0242, (01) 0 9342395 00015 1 ISG0370, (01) 0 9342395 00024 3 IPG0539, (01) 0 9342395 00009 0 BG1190

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US nationwide/ Worldwide Distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 5, 2023
Recall initiation date
Mar 16, 2023
Center classification date
Jun 27, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.