Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-... — FDA record Z-1950-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 23, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida" CI M and Sky CI" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implant battery (P675). Regular 675 hearing instrument batteries will not be powerful enough for a cochlear implant. The suggested manufacturer and battery is PowerOne P675 Implant Plus battery. The M Zn-Air Battery Pak is available with non-tamperproof and tamperproof cartridges. Component: Not Applicable
Reason
The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.
Classification
Class II
Recalling firm
Advanced Bionics, LLC
Quantity
45173
Agency-reported status
Ongoing
Source record identifier
Z-1950-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: All lots manufactured up to February 3, 2026. Refer to Exhibit_0006

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, China, Chile, Croatia, Cuba, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kazakhstan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Trinidad and Tobago, United Arab Emirates, and United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 6, 2026
Recall initiation date
Mar 23, 2026
Center classification date
Apr 27, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records