Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-... — FDA record Z-1946-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 30, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
Reason
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Classification
Class II
Recalling firm
Zimmer, Inc.
Quantity
66960
Agency-reported status
Ongoing
Source record identifier
Z-1946-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301014(10)85390029 ; Individual UDI (01)00889024376564(17)301014(10)85390029 ; Lot Number 85390029 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301103(10)85682882 ; Individual UDI (01)00889024376564(17)301103(10)85682882 ; Lot Number 85682882 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301001(10)85390028 ; Individual UDI (01)00889024376564(17)301001(10)85390028 ; Lot Number 85390028 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300422(10)82395239 ; Individual UDI (01)00889024376564(17)300422(10)82395239 ; Lot Number 82395239 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300423(10)82395240 ; Individual UDI (01)00889024376564(17)300423(10)82395240 ; Lot Number 82395240 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300929(10)85390027 ; Individual UDI (01)00889024376564(17)300929(10)85390027 ; Lot Number 85390027 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300924(10)85390026 ; Individual UDI (01)00889024376564(17)300924(10)85390026 ; Lot Number 85390026 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300415(10)82395238 ; Individual UDI (01)00889024376564(17)300415(10)82395238 ; Lot Number 82395238 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300922(10)85390025 ; Individual UDI (01)00889024376564(17)300922(10)85390025 ; Lot Number 85390025 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300826(10)85390022 ; Individual UDI (01)00889024376564(17)300826(10)85390022 ; Lot Number 85390022 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300918(10)85390024 ; Individual UDI (01)00889024376564(17)300918(10)85390024 ; Lot Number 85390024 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300722(10)85390023 ; Individual UDI (01)00889024376564(17)300722(10)85390023 ; Lot Number 85390023 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)310122(10)86970970 ; Individual UDI (01)00889024376564(17)310122(10)86970970 ; Lot Number 86970970

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MS, NC, NE, NM, NY, OH, OK, PA, SD, TN, TX, UT, WA, WI and the countries of Canada, Australia, New Zealand, Malaysia, EMEA, Chile.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 6, 2026
Recall initiation date
Mar 30, 2026
Center classification date
Apr 24, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.