Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5,... — FDA record Z-1943-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Arial Emergency and Nurse Call System.
Reason
The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.
Classification
Class II
Recalling firm
Securitas Healthcare LLC
Quantity
661 sites with mobile application user licenses
Agency-reported status
Ongoing
Source record identifier
Z-1943-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityArial Mobile v11.0.0 UDI: (01)00840948100963(10)1100.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Canada, England, Hong Kong, France, Qatar, and Spain.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 12, 2024
Recall initiation date
Apr 11, 2024
Center classification date
May 31, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.