Recall archive / FDA device enforcement
ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample) — FDA record Z-1910-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date May 2, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)
- Reason
- During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.
- Classification
- Class II
- Recalling firm
- Healthmark Industries Co., Inc.
- Quantity
- 193035 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1910-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Lot Numbers: 200014 200015 200016 900153 900154 900155 900156 900157 900158 900159 900160 900161 900162 900163 900164 900165 900166 900167 900168 900169 900170 900171 900172 900173 900174 900175 900176 900177 103564 103312 103293 102807 102096 101673 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide and Internationally to countries of: Canada, Malaysia.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 11, 2025
- Recall initiation date
- May 2, 2025
- Center classification date
- Jun 5, 2025