Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010 — FDA record Z-1909-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 2, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010
Reason
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.
Classification
Class II
Recalling firm
Healthmark Industries Co., Inc.
Quantity
250000 units
Agency-reported status
Ongoing
Source record identifier
Z-1909-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Numbers: 200001 200002 200003 200004 200005 200006 200007 200008 200009 200010 200011 200012 200013

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide and Internationally to countries of: Canada, Malaysia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 11, 2025
Recall initiation date
May 2, 2025
Center classification date
Jun 5, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.