Access Testosterone assay, Catalog Number 33560 — FDA record Z-1901-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Access Testosterone assay, Catalog Number 33560
Reason
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
Classification
Class II
Recalling firm
Beckman Coulter, Inc.
Quantity
70 units
Agency-reported status
Ongoing
Source record identifier
Z-1901-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 15099590231347; Serial Number: 538081

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US State: FL. UAE, Vietnam

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 11, 2025
Recall initiation date
May 5, 2025
Center classification date
Jun 2, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.