Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VI Software... — FDA record Z-1900-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 24, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: N/A
Reason
The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.
Classification
Class II
Recalling firm
Lucid Diagnostics, Inc.
Quantity
220 units (US)
Agency-reported status
Ongoing
Source record identifier
Z-1900-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Product sent out to clinical site: Part #: EC25T10VI ; GTIN Number: 00860007510977 ; Lot #'s: 483 (Qty: 120), 502 (Qty: 100) Product at Distributor in Quarantine: Part #: EC25T10VI ; GTIN Number: 00860007510977 ; Lot #'s: 539 (Qty: 99)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of MN, MD.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 5, 2024
Recall initiation date
Apr 24, 2024
Center classification date
May 24, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.