Captus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests. — FDA record Z-1892-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 22, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Captus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
Reason
Complaints of unexpected detachment of the collimator have been reported.
Classification
Class II
Recalling firm
Mirion Technologies (Capintec), Inc.
Quantity
1,275 units
Agency-reported status
Ongoing
Source record identifier
Z-1892-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Model No: 5430-30151; UDI 859942006096, S/N range 940001 through 941398.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, FRANCE, GERMANY, GUAM, HONDURAS, HUNGARY, INDIA, ISRAEL, ISREAL, ITALY, JAPAN, KOREA, LATVIA, MALAYSIA, NETHERLANDS, NEW ZEALAND, PAKISTAN, PHILIPPINES, PHILIPPINES , PHILLIPINES, PUERTO RICO, QATAR, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TAIWAN , THAILAND, TURKEY, UNITED KINGDOM, VIETNAM.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 5, 2024
Recall initiation date
Apr 22, 2024
Center classification date
May 24, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records