Recall archive / FDA device enforcement
Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IP... — FDA record Z-1890-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Apr 18, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
- Reason
- Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
- Classification
- Class II
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Quantity
- 14,977
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1890-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 through 26-MAR-2026; M365DB12320/8714729985051/100104 - 753200/09-OCT-2020 through 21-MAR-2026 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Italy, Jordan, Kazakhstan, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Australia, Thailand, Singapore, Hong Kong, Korea, India, Japan.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 5, 2024
- Recall initiation date
- Apr 18, 2024
- Center classification date
- May 24, 2024