Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24 — FDA record Z-1884-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
Reason
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Classification
Class II
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
418 units
Agency-reported status
Ongoing
Source record identifier
Z-1884-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 29, 2024
Recall initiation date
Apr 11, 2024
Center classification date
May 22, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.