Cascade IOMAX Cortical Module, REF: 190296-200 — FDA record Z-1882-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 23, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Cascade IOMAX Cortical Module, REF: 190296-200
Reason
There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.
Classification
Class II
Recalling firm
Cadwell Industries Inc
Quantity
54 units
Agency-reported status
Ongoing
Source record identifier
Z-1882-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00840067100653/ Serial numbers: 19029603AA1223002 19029603AA1223003 19029603AA1223004 19029603AA1223005 19029603AA1223006 19029603AA1223007 19029603AA1223011 19029603AA1223012 19029603AA1223013 19029603AA1223014 19029603AA1223015 19029603AA1223016 19029603AA1223017 19029603AA1223018 19029603AA1223019 19029603AA1223020 19029603AA1223024 19029603AB1223021 19029603AB1223023 19029603AA1223025 19029603AA0124001 19029603AA0124002 19029603AA0224008 19029603AA0124007 19029603AA0124012 19029603AA0124017 19029603AA0124004 19029603AA0124010 19029603AA0124016 19029603AA0224010 19029603AA0124015 19029603AA1223026 19029603AA0124008 19029603AA0124009 19029603AA1223027 19029603AA1223028 19029603AA1223029 19029603AA0124003 19029603AA0224001 19029603AA1223030 19029603AA1223008 19029603AA1223009 19029603AA0124013 19029603AA1223031 19029603AA1223001 19029603AA0224002 19029603AA0224003 19029603AA0224004 19029603AA0224005 19029603AA0224006 19029603AA0224007 19029603AA0124014 19029603AA0224009 19029603AA0124011

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 29, 2024
Recall initiation date
Apr 23, 2024
Center classification date
May 22, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.