Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF D... — FDA record Z-1876-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5 mm ID; (3) REF DYNJ18870, Polyurethane Cuff 7.0 mm ID; (4) REF DYNJ18875, Polyurethane Cuff 7.5 mm ID; (5) REF DYNJ18880, Polyurethane Cuff 8.0 mm ID; and (6) REF DYNJ18885, Polyurethane Cuff 8.5 mm ID. Medline SubG Endotracheal Tube with Subglottic Suctioning packaged into the following convenience kits: (1) Medline Intubation Kit - Drawer 3, Reorder No. ACC010502, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (2) Medline Intubation Tube - Drawer 3, Reorder No. ACC010527, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (3) Medline Adult Intubation, Reorder No. ACC010540, containing 8.5 size subglottic cuff, 3 kits/case; (4) Medline Adult Intubation, Reorder No. ACC010540A, containing 8.5 size subglottic cuff, 3 kits/case; (5) Medline Airway Kit - Drawer 3, Reorder No. ACC010717, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 3 kits/case; (6) Medline Airway Kit, Reorder No. DYNDJ1132, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 26 kits/case; (7) Medline Adult Intubation W Evac, Reorder No. DYNDJ1133, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 10 kits/case; (8) Medline Backup Kit, Reorder No. DYNJ909501, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (9) Medline Backup Kit, Reorder No. DYNJ909501A, containing 6.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (10) Medline Adult Intubation Tray, Reorder No. DYNJAA269, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 10 kits/case.
- Reason
- Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.
- Classification
- Class I
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 168,632 tubes and 13,092 kits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1876-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI numbers for the individual tubes: (1) DYNJ18860 - UDI-DI 20888277652672; (2) DYNJ18865 - UDI-DI 20888277652689; (3) DYNJ18870 - UDI-DI 20888277652696; (4) DYNJ18875 - UDI-DI 20888277652702; (5) DYNJ18880 - UDI-DI 20888277652719; (6) DYNJ18885 - UDI-DI 20888277657318. Kit code information: (1) Kit ACC010502 - Kit UDI-DI: 10193489846171; Case UDI-DI: 40193489846172; lot numbers: 23FDC233, 23FDB211, 23EDC153, 23EDB775, 23DDB280, 23CDB943, 23CDA556, 22LDA029, 22IDA550, 22HDA813, 22FDC033, 22CDC284, 21JDC054, 21IDB345, 21HDB405, and 21DDB305; (2) Kit ACC010527 - Kit UDI-DI: 10193489884333; Case UDI-DI: 40193489884334; lot numbers: 23JDA443, 23IDA363, 23IDA232, 23HDC331, 23HDA117, 23GDB600, 23FDC372, 23FDB913, 23FDA966, 23FDA310, 23EDC371, 23EDC154, 23EDC094, 23EDB732, 23EDA387, 23CDB838, 23CDA557, 23BDC008, 23BDA945, 23ADB791, 22LDA005, 22JDB800, 22HDB976, 22HDA451, 22GDB041, 21LDB937, 21LDB416, 21JDB262, 21IDB491, 21IDA193, 21GDB207, and 21FDC754; (3) Kit ACC10540 - Kit UDI-DI: 10193489977547; Case UDI-DI 40193489977548; lot number: 23ADB410; (4) Kit ACC10540A - Kit UDI-DI 10195327330088; Case UDI-DI 40195327330089; lot numbers: 23HDA588, 23GDB107, and 23BDA684; (5) Kit ACC010717 - Kit UDI-DI 10195327458997; Case UDI-DI 40195327458998; lot numbers: 24ADA577, 23LDA731, and 23IDA629; (6) Kit DYNDJ1132 - Kit UDI-DI 10195327454821; Case UDI-DI 40195327454822; lot numbers: 23LDA733, 23LDA617, 23JDC021, and 23HDA569; (7) Kit DYNDJ1133 - Kit UDI-DI 10195327428532; Case UDI-DI 40195327428533; lot numbers: 24BMB124, 23LMB682, 23HMF259, and 23HMD765; (8) Kit DYNJ909501 - Kit UDI-DI 10195327384852; Case UDI-DI 40195327384853; lot number: 23GBI504; (9) Kit DYNJ909501A - Kit UDI-DI 10195327549817; Case UDI-DI 40195327549818; lot numbers: 24BBD129, 24ABK281, and 23KBR773; (10) Kit DYNJAA269 - Kit UDI-DI 10195327209049; Case UDI-DI 40195327209040; lot numbers: 24BBE224, 24ABM718, 24ABD835, 23LBB055, 23JBP694, 23JBQ783, 23IBU821, 23IBF348, 23HBP188, 23CBX318, 23CBX319, 23BBR494, 23ABI514,22LBI988, 22KBH009, and 22JBH478. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 5, 2024
- Recall initiation date
- Feb 26, 2024
- Center classification date
- May 28, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.