NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454,... — FDA record Z-1864-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 28, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
Reason
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
Classification
Class II
Recalling firm
NxStage Medical Inc
Quantity
115,465 cases (230,930 units)
Agency-reported status
Ongoing
Source record identifier
Z-1864-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitya. UDI-DI number: M535RFP4000. Device Catalog Number RFP-400. Lot Numbers: Q2407307, Q2407308, Q2407309. b. UDI-DI number: M535RFP4010. Device Catalog Number RFP-401. Lot Numbers: Q2407061, Q2407147, Q2407148, Q2407149, Q2407313, Q2407386, Q2407387, Q2407388, Q2407389, Q2407390. c. UDI-DI number: M535RFP4020. Device Catalog Number RFP-402. Lot Numbers: Q2407213, Q2407214. d. UDI-DI number: M535RFP4030. Device Catalog Number RFP-403. Lot Number: Q2407311. e. UDI-DI number: M535RFP4040. Device Catalog Number RFP-404. Lot Numbers: Q2407062, Q2407063, Q2407064, Q2407065, Q2407067, Q2407068, Q2407069, Q2407070, Q2407071, Q2407217, Q2407218, Q2407219, Q2407220, Q2407221, Q2407222, Q2407223, Q2407224, Q2407225, Q2407301, Q2407302, Q2407303, Q2407304, Q2407305, Q2407306, Q2407391, Q2407392, Q2407405, Q2407406. f. UDI-DI number: M535RFP4060. Device Catalog Number RFP-406. Lot Numbers: Q2407072, Q2407137, Q2407140, Q2407141, Q2407216, Q2407310, Q2407312. g. UDI-DI number: M535RFP4530. Device Catalog Number RFP-453. Lot Number: Q2407066. h. UDI-DI number: M535RFP4540. Device Catalog Number RFP-454. Lot Numbers: Q2407145, Q2407146, Q2407226. j. UDI-DI number: M535RFP4560. Device Catalog Number RFP-456. Lot Numbers: Q2407138, Q2407139, Q2407142, Q2407143, Q2407144, Q2407227.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 4, 2025
Recall initiation date
Apr 28, 2025
Center classification date
May 28, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.