Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0... — FDA record Z-1858-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 22, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0026), Catalyst PT (SP001-0027), Catalyst X4 (SP002-0035), Catalyst+ (SP003-0001), Catalyst+ HD (SP003-0002), Catalyst+ PT (SP003-0008), Catalyst+ X4, (SPOO3-0009) Product Description: Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not a component
Reason
Software issue with scanning equipment that can results in the filed rotation not being applied correctly.
Classification
Class II
Recalling firm
C-RAD POSITIONING AB
Quantity
44 units
Agency-reported status
Ongoing
Source record identifier
Z-1858-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll Catalyst products using software Version: c4D 6.5.1 SP2 Catalyst -SP001-0025/UDI: 07350005280024; Catalyst HD- SP001-0026/UDI: 07350005280024; Catalyst PT- SP001-0027/UDI: 07350005280024; Catalyst X4- SP002-0035/UDI: 07350005280024; Catalyst+ - REF: SP003-0001/UDI: 17350005280762 Catalyst+ HD- SP003-0002/UDI: 17350005280779 Catalyst+ PT- SP003-0008/UDI: 17350005280793 Catalyst+ X4 - SP003-0009/UDI: 17350005280809

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AK, FL, WA, VA, WA, NC, IL, AR, NC and the countries of Germany, Italy, Uzbekistan, South Korea, Vietnam, Austria, Australia, China, Estonia, Belgium, Uzbekistan.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 4, 2025
Recall initiation date
Apr 22, 2025
Center classification date
May 28, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records