Intera 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device.... — FDA record Z-1853-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 3, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Intera 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781295
Reason
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Classification
Class II
Recalling firm
Philips North America Llc
Quantity
200 units
Agency-reported status
Ongoing
Source record identifier
Z-1853-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: N/A Serial Number: 30199 30200 30201 30062 30191 30240 30241 30242 30305 30315 30125 30088 30113 30222 30252 30268 30028 30038 30058 30069 30081 30095 30100 30117 30118 30138 30164 30167 30168 30169 30170 30192 30197 30207 30209 30217 30228 30246 30248 30249 30256 30262 30267 30281 30284 30289 30297 30298 30306 30307 30312 30314 30210 30219 30245 30254 30276 30029 30043 30052 30061 30083 30101 30107 30114 30115 30116 30119 30120 30122 30130 30137 30140 30153 30154 30158 30159 30160 30163 30173 30174 30177 30178 30179 30196 30203 30251 30253 30259 30265 30266 30277 30279 30280 30282 30285 30291 30293 30104 30105 30216 30244 30290 30308 30316 30070 30208 30102 30112 30133 30206 30214 30224 30238 30261 30264 30274 30295 30303 30051 30084 30149 30215 30227 30232 30233 30234 30235 30035 30055 30143 30148 30272 30288 30294 8281 30304 30142 30218 30109 30131 30063 30091 30094 30202 30059 30099 30175 30176 30198 30237 30263 30269 30270 30302 30309 8002 8143 8122 8855 8404 30053 30075 30145 30064 30065 30123 30171 30085 30190 30226 30230 30231 30213 30135 30152 30155 30180 30181 30182 30183 30184 30185 30186 30187 30189 30220 30221 30223 30243 30250 30283 30310 30311 30017 30030 30047 30098 10193

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 29, 2024
Recall initiation date
May 3, 2024
Center classification date
May 17, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records