ABBOTT POINT OF CARE TEST CREA I-STAT (25/BX) ABBTPT CARTRIDGE. — FDA record Z-1846-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 24, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ABBOTT POINT OF CARE TEST CREA I-STAT (25/BX) ABBTPT CARTRIDGE.
Reason
transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.
Classification
Class II
Recalling firm
Mckesson Medical-Surgical Inc. Corporate Office
Quantity
25 units
Agency-reported status
Ongoing
Source record identifier
Z-1846-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog No. 03P8425; GTIN: 10054749000125; Order No. 46085650.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AL, AR, IL, KS, MT, NC, NJ, OR, SC, TN & WV.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 4, 2025
Recall initiation date
Apr 24, 2025
Center classification date
May 27, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.