Roadrunner PC Wire Guide (Nimble), Guidewire Reference Part Number/GPN: RFSPC-35-145 G09607 — FDA record Z-1838-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 16, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Roadrunner PC Wire Guide (Nimble), Guidewire Reference Part Number/GPN: RFSPC-35-145 G09607
Reason
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Classification
Class II
Recalling firm
Cook Incorporated
Quantity
277 US; 138 OUS
Agency-reported status
Ongoing
Source record identifier
Z-1838-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityRFSPC-35-145 G09607 UDI-DI: 00827002096073 Lot/Expiration Date: 14831437 07-07-2025 14831438 07-07-2025 14835040 11-07-2025 14840287 13-07-2025 14840288 13-07-2025 14840309 13-07-2025

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 5, 2022
Recall initiation date
Aug 16, 2022
Center classification date
Sep 28, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records