Recall archive / FDA device enforcement
Fixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476 — FDA record Z-1832-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 16, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Fixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476
- Reason
- Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
- Classification
- Class II
- Recalling firm
- Cook Incorporated
- Quantity
- 58 US; 5 OUS
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1832-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | THSCF-25-180-3 G02165 UDI-DI: 00827002021655 Lots/Expiration Dates: 14797761 16-06-2027 14813704 27-06-2027 TSCF-25-80-3 G00476 UDI-DI: 00827002004764 Lots/Expiration Dates: NS14797766 16-06-2027 NS14807600 22-06-2027 NS14821727 30-06-2027 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Oct 5, 2022
- Recall initiation date
- Aug 16, 2022
- Center classification date
- Sep 28, 2022