Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software ... — FDA record Z-1822-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 16, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable
Reason
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
Classification
Class II
Recalling firm
THERAKOS DEVELOPMENT LIMITED
Quantity
11,499 units
Agency-reported status
Ongoing
Source record identifier
Z-1822-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel/Catalog Number: CLXUSA; UDI-DI: 20705030200003; Lot Numbers: N301, N339, N340, N341, N342, N343, N344, N345, N346, N347; P302, P303, P304, P305, P306, P307, P308, P309, P310, P311, P312, P313, P314, P319, P320.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 22, 2026
Recall initiation date
May 16, 2025
Center classification date
Apr 16, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.