Novalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respirator... — FDA record Z-1822-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 8, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Novalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation and CO2 removal) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure Product Code: F30000163
Reason
Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during use of the Novalung system.
Classification
Class II
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
88 units
Agency-reported status
Ongoing
Source record identifier
Z-1822-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 4057224 (+ serial number) Serial Numbers:XCONUS0001 to XCONUS0124

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 22, 2024
Recall initiation date
Apr 8, 2024
Center classification date
May 15, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.