Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295). The Mar... — FDA record Z-1822-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 12, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295). The Mark I model is no longer produced, but the legacy humidifier bottle component design can also be used with the BT37 Mark II Benchtop Incubator.
Reason
There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.
Classification
Class II
Recalling firm
CooperSurgical, Inc.
Quantity
5,182 kits
Agency-reported status
Ongoing
Source record identifier
Z-1822-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Numbers 05593V291121 18-1192 1812-68 18-1268 G000575 G003811 G003893 G004517 G004874 G005109 G005476 G005723 E160415 Expiration date range: 27 Jun 2022 to 07 Dec 2024

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS MA MI MN MO NJ NY OH OR SC TX VA VT WA WI WV Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 5, 2022
Recall initiation date
Aug 12, 2022
Center classification date
Sep 28, 2022