Apo B Reagent, REF: OSR6143 — FDA record Z-1815-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 23, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Apo B Reagent, REF: OSR6143
Reason
The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.
Classification
Class III
Recalling firm
Beckman Coulter Ireland, Inc.
Quantity
2747 units
Agency-reported status
Ongoing
Source record identifier
Z-1815-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot # 2631/UDI: 15099590010409

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 4, 2025
Recall initiation date
Apr 23, 2025
Center classification date
May 23, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records