Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F — FDA record Z-1814-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 1, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F
Reason
Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.
Classification
Class I
Recalling firm
SunMed Holdings, LLC
Quantity
11,358
Agency-reported status
Ongoing
Source record identifier
Z-1814-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: EA: 10884389164822 CS: 40884389164823 Lots: 526782 526790 526796 526797 526798 526800 526802 526804 526805 526806 526807 526808 526809 526810 526814 526815 526816 526811 526817 526818

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution via Medline.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 11, 2025
Recall initiation date
May 1, 2025
Center classification date
May 30, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.