Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7 — FDA record Z-1797-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 12, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7
Reason
Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.
Classification
Class I
Recalling firm
Insulet Corporation
Quantity
1,240,115 units
Agency-reported status
Ongoing
Source record identifier
Z-1797-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7. UDI-DI: 10385083000527. Lot Codes: PH1U02252541, PH1U03282511, PH1U03282522, PH1U03312511, PH1U03312521, PH1U04012511, PH1U04012521, PH1U05052511, PH1U08162531, PH1U08182531, PH1U08182541, PH1U09242511, PH1U09242521, PH1U09242531, PH1U09252521, PH1U09252531, PH1U10152541, PH1U10162531, PH1U10162541, PH1U10172531, PH1U10172541, PH1U10182531, PH1U10182541, PH1U10202511, PH1U10202521, PH1U10202531, PH1U10202541, PH1U10212531, PH1U10212541, PH1U10222531, PH1U10222541, PH1U10232531, PH1U10232541, PH1U10242521, PH1U10242531, PH1U10242541.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 22, 2026
Recall initiation date
Mar 12, 2026
Center classification date
Apr 10, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records