Recall archive / FDA device enforcement
COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Refere... — FDA record Z-1787-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 5, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
- Reason
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Classification
- Class II
- Recalling firm
- Cook Incorporated
- Quantity
- 70 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1787-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Reference Part Number C-NTSE-2.4-115-NC3, UDI (01)00827002362505(17)261103(10)15741134, Lot Number 15741134; Reference Part Number C-NTSE-2.4-115-NCT4, UDI (01)00827002362512(17)270118(10)15855141, Lot Number 15855141. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Apr 15, 2026
- Recall initiation date
- Mar 5, 2026
- Center classification date
- Apr 6, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.