Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule an... — FDA record Z-1786-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 5, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115
Reason
Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
Classification
Class III
Recalling firm
Randox Laboratories Ltd.
Quantity
40 units
Agency-reported status
Ongoing
Source record identifier
Z-1786-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 05055273214970 Serial Numbers: 052-23-0424 052-23-0425 052-23-0434 052-23-0437 052-23-0618 052-23-0611 052-23-0516 052-23-0522 052-23-0490 052-23-0491 052-23-0439 052-23-0440 052-23-0441 052-23-0445 052-23-0449 052-23-0450 052-23-0451 052-23-0452 052-23-0438 052-22-0276 052-22-0277 052-22-0279 052-23-0416 052-22-0283 052-22-0271 052-21-0260 052-21-0255 052-21-0249 052-23-0418 052-23-0419 052-23-0415 052-23-0417 052-22-0268 052-23-0420 052-22-0284 052-23-0448 052-23-0444 052-23-0443 052-23-0442 052-23-0609

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 15, 2024
Recall initiation date
Apr 5, 2024
Center classification date
May 9, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.