Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19
- Reason
- Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
- Classification
- Class II
- Recalling firm
- Micro-X Ltd.
- Quantity
- 42 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1786-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model Number: MXU-RV19 UDI Codes: (01)09357123000013(11)210326(21)00222 (01)09357123000013(11)210628(21)00234 (01)09357123000013(11)210629(21)00235 (01)09357123000013(11)210629(21)00236 (01)09357123000013(11)210630(21)00237 (01)09357123000013(11)210708(21)00238 (01)09357123000013(11)210721(21)00240 (01)09357123000013(11)210910(21)00261 (01)09357123000013(11)210913(21)00262 (01)09357123000013(11)210914(21)00263 (01)09357123000013(11)210921(21)00264 (01)09357123000013(11)210922(21)00265 (01)09357123000013(11)210923(21)00266 (01)09357123000013(11)210927(21)00267 (01)09357123000013(11)211006(21)00268 (01)09357123000013(11)211008(21)00269 (01)09357123000013(11)211011(21)00270 (01)09357123000013(11)211011(21)00271 (01)09357123000013(11)211012(21)00272 (01)09357123000013(11)211013(21)00273 (01)09357123000013(11)211013(21)00274 (01)09357123000013(11)211210(21)00288 (01)09357123000013(11)211213(21)00289 (01)09357123000013(11)211213(21)00290 (01)09357123000013(11)220329(21)00305 (01)09357123000013(11)220330(21)00306 (01)09357123000013(11)220331(21)00307 (01)09357123000013(11)220401(21)00308 (01)09357123000013(11)220404(21)00309 (01)09357123000013(11)220405(21)00310 (01)09357123000013(11)220406(21)00311 (01)09357123000013(11)220404(21)00312 Serial Numbers: 222 234 235 236 237 238 240 261 262 263 264 265 266 267 268 269 270 271 272 273 274 288 289 290 305 306 307 308 309 310 311 312 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Oct 5, 2022
- Recall initiation date
- Aug 31, 2022
- Center classification date
- Sep 23, 2022