Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid ... — FDA record Z-1776-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.
Reason
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Classification
Class I
Recalling firm
Medtronic Navigation, Inc.
Quantity
3286 applications.
Agency-reported status
Ongoing
Source record identifier
Z-1776-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDIs: 00763000234867, 00763000125295, 00763000063177, 00643169865341/ Software Versions: 1.2.0, 1.1.0, 1.0.3, 1.0.2, and 1.0.1.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Republic Of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 15, 2024
Recall initiation date
Mar 26, 2024
Center classification date
May 9, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.