What the record says

Product
The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.
Reason
***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg
Classification
Class II
Recalling firm
Randox Laboratories Ltd.
Quantity
2 units
Agency-reported status
Terminated
Source record identifier
Z-1772-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of WV, NC.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 28, 2022
Recall initiation date
Aug 1, 2022
Center classification date
Sep 19, 2022
Termination date
Dec 13, 2024