What the record says

Product
AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US
Reason
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
Classification
Class II
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
24 units
Agency-reported status
Ongoing
Source record identifier
Z-1769-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: (01)4251285200234(11)210628(21)1CHS5379 (01)4251285200234(11)160927(21)6CHS1270 (01)4251285200234(11)161109(21)6CHS1394 (01)4251285200234(11)161109(21)6CHS1435 (01)4251285200234(11)170131(21)7CHS1569 (01)4251285200234(11)170201(21)7CHS1583 (01)4251285200234(11)170213(21)7CHS1619 (01)4251285200234(11)170314(21)7CHS1641 (01)4251285200234(11)170519(21)7CHS1790 (01)4251285200234(11)170807(21)7CHS1953 (01)4251285200234(11)180326(21)8CHS2197 (01)4251285200234(11)181010(21)8CHS2608 (01)4251285200234(11)181102(21)8CHS2633 (01)4251285200234(11)181107(21)8CHS2644 Serial Numbers: 1CHS5379 6CHS1192 6CHS1194 6CHS1195 6CHS1196 6CHS1197 6CHS1198 6CHS1199 6CHS1200 6CHS1201 6CHS1233 6CHS1270 6CHS1394 6CHS1435 7CHS1569 7CHS1583 7CHS1619 7CHS1641 7CHS1790 7CHS1953 8CHS2197 8CHS2608 8CHS2633 8CHS2644

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 28, 2022
Recall initiation date
Jun 21, 2022
Center classification date
Sep 19, 2022