ORCA FOAM STANDARD CUBE 10x10x10 mm, Absorbable Gelatin Sponge Hemostat, USP, REF ORCA FOAM — FDA record Z-1768-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 27, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ORCA FOAM STANDARD CUBE 10x10x10 mm, Absorbable Gelatin Sponge Hemostat, USP, REF ORCA FOAM
Reason
In routine bioburden testing, higher than acceptable levels of Endotoxins were observed.
Classification
Class II
Recalling firm
GELITA MEDICAL GMBH
Quantity
3596 boxes of 50 devices
Agency-reported status
Terminated
Source record identifier
Z-1768-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 00860003109915, Batch Numbers: T02371/1, expiration date 10.02.2026; T02371/2, expiration date 10.02.2026; T02395/1, expiration date 07.05.2026; T02394/2, expiration date 07.05.2026

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of Alabama.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 28, 2022
Recall initiation date
Jul 27, 2022
Center classification date
Sep 19, 2022
Termination date
Sep 22, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.