Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 3... — FDA record Z-1759-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 5, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.
Reason
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
Classification
Class II
Recalling firm
Randox Laboratories Ltd.
Quantity
80,366 units (US: 242 units; OUS: 80,094 units)
Agency-reported status
Ongoing
Source record identifier
Z-1759-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalogue No. CAL2351 & CAL10388; GTIN: 05055273200966; Lot No. 1260UE, 1262UE, 1295UE, 1297UE, 1298UE, 1315UE, 1325UE, 1326UE, 1295UE, & 1325UE.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic: Nationwide Distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 15, 2024
Recall initiation date
Mar 5, 2024
Center classification date
May 3, 2024